Camera Calibration Technical Report:
Calibration
|
Next: Image Correction
Up: Methods
Previous: Hardware
To use the calibration method outlined above requires the determination
of the graph of G(I) against G(I'). This can be obtained
using a single neutral density filter (possibly unknown value), plus
any microscope slide with a reasonable variation of densities.
We assume that each of the elements of the
array has a transfer function of the form
gi(I) = ai + f(bi I),
where i is an element index
and each may have a different
dark-field value, ai and a different shading, bi.
Variation in shading coefficient bi arises from non-even illumination
and from differences in individual CCD array elements for example their
respective quantum efficiencies.
The procedure to generate the
points,
(G(I), G(I')), is:
- 1.
- digitise an entire dark field image by
switching the microscope lamp off;
- 2.
- select a field on a slide that exhibits a reasonable
range of densities and adjust the lamp level so that the brighter
parts of the field almost saturate the digitised image;
- 3.
- digitise the field twice, without moving the slide, first
without and then with the neutral density filter in the illumination path,
1
typically before the condenser. Subtract the dark-field image
from each of these new images;
- 4.
- accumulate the values
(G(I), G(I')) by scanning through
the two images, corresponding pixels from the two images are the
required value pairs, G(I') is the value from the image
with the neutral density filter and G(I) is the value from the
image without the filter 2;
- 5.
- inspect the range of values extracted from the images;
if parts of the response range are missing then adjust the
lamp voltage and repeat from 3;
- 6.
- establish the values
;
- 7.
- select a starting value G0 near the bright end of the range to
correspond to the arbitrary offset density
and tabulate the function
by the method outlined
earlier.
Next: Image Correction
Up: Methods
Previous: Hardware
Richard Baldock
1998-06-16