|
Camera Calibration Technical Report:
Image Correction |
The calibration procedure results in a conversion table from measured value to a density. To determine the density values in a particular image requires a corresponding dark-field (zero light) and bright-field (no object) image. Since the dark-field image clearly does not depend on the microscope optics or illumination, the dark-field image captured during calibration may be used for subsequent measurements. However the bright-field image must be re-captured whenever the slide, objective, condenser or lamp brightness are adjusted. The bright-field image is obtained as follows:
The bright-field may be used to correct subsequent images from the same slide
as long as the microscope settings are not adjusted. To correct
an image subtract the dark-field value,
convert each pixel value to a density and
subtract the bright-field density: